  • Title: Optimization of an algorithm for measurements of velocity vector components using a three-wire sensor.
    Author: Ligeza P, Socha K.
    Journal: Rev Sci Instrum; 2007 Oct; 78(10):105104. PubMed ID: 17979455.
    Abstract:
    Hot-wire measurements of velocity vector components use a sensor with three orthogonal wires, taking advantage of an anisotropic effect of wire sensitivity. The sensor is connected to a three-channel anemometric circuit and a data acquisition and processing system. Velocity vector components are obtained from measurement signals, using a modified algorithm for measuring velocity vector components enabling the minimization of measurement errors described in this paper. The standard deviation of the relative error was significantly reduced in comparison with the classical algorithm.