These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure during spinal anaesthesia for caesarean section: a preliminary descriptive study.
    Author: Ngan Kee WD, Tam YH, Khaw KS, Ng FF, Critchley LA, Karmakar MK.
    Journal: Anaesthesia; 2007 Dec; 62(12):1251-6. PubMed ID: 17991262.
    Abstract:
    We describe the novel use of a closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure in 53 patients having spinal anaesthesia for elective caesarean section. A simple on-off algorithm was used that activated an intravenous phenylephrine infusion at 100 microg.min(-1) when systolic blood pressure was less than or equal to baseline and stopped the infusion when systolic blood pressure exceeded baseline. Up to uterine incision, 94.6% of all systolic blood pressure measurements were within the range (baseline +/- 20%). Seven patients (13.2%) had one or more episodes of hypotension (systolic blood pressure < 80% of baseline) and 23 patients (37.7%) had one or more episodes of hypertension (systolic blood pressure > 120% of baseline). No patient had nausea or vomiting and in no case was umbilical arterial blood pH < 7.2. Calculated system performance parameters were comparable with those of previously published closed-loop systems and provide a reference for the potential development and comparison of more advanced algorithms.
    [Abstract] [Full Text] [Related] [New Search]