These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A sub-microwatt low-noise amplifier for neural recording. Author: Holleman J, Otis B. Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2007; 2007():3930-3. PubMed ID: 18002859. Abstract: In this paper we present a pre-amplifier designed for neural recording applications. Extremely low power dissipation is achieved by operating in an open-loop configuration, restricting the circuit to a single current branch, and reusing current to improve noise performance. Our amplifier exhibits 3.5 microVrms of input-referred noise and has a digitally-controlled gain between 36 dB and 44 dB. The amplifier is AC-coupled, with a pass-band from 0.3 Hz to 4.7 kHz. The circuit is implemented in a 0.5 microm SOI Bi-CMOS process and consumes 805nA from a 1.0V supply, corresponding to a noise efficiency factor (NEF) of 1.8, which is the lowest reported NEF to date.[Abstract] [Full Text] [Related] [New Search]