These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Analysis of CoQ10 in rat serum by ultra-performance liquid chromatography mass spectrometry after oral administration.
    Author: Li L, Pabbisetty D, Carvalho P, Avery MA, Avery BA.
    Journal: J Pharm Biomed Anal; 2008 Jan 07; 46(1):137-42. PubMed ID: 18054195.
    Abstract:
    A UPLC-MS method for determining Coenzyme Q(10) (CoQ(10)) levels in rat serum was developed. CoQ(10) was quantitatively extracted into 2-propanol using a fast extraction procedure. The separation of CoQ(10) was performed on a Waters Acquity UPLCtrade mark BEH C(18) column (1.7 microm, 1.0 mm x 50 mm) with the mobile phase containing acetonitrile, 2-propanol, and formic acid (90:10:0.1) over 5 min. The sensitivity of this method allows for the quantitation of 50 ng/mL CoQ(10) in serum (S/N=10). The linearity of this method was found to be from 50 to 20,000 ng/mL. The precision was less than 10% (intra- and inter-day), and the average extraction recovery was between 90 and 105%. This procedure provides a precise, sensitive and direct assay method for the determination of CoQ(10) in rat serum after oral administration. This method could be applied to further pharmacokinetic studies of CoQ(10).
    [Abstract] [Full Text] [Related] [New Search]