  • Title: Optimization of the intensity gain of multiple-focus phased-array heating patterns.
    Author: Ebbini ES, Cain CA.
    Journal: Int J Hyperthermia; 1991; 7(6):953-73. PubMed ID: 1806647.
    Abstract:
    A new technique for enhancing the intensity gain at the focal points in multiple-focus patterns is introduced. The new technique is shown to be effective in reducing the interference typically associated with multiple-focus patterns. This reduction in interference patterns allows multiple-focus scanning to generate highly localized heating. Simulation results indicate that multiple-focus scanning not only provides an alternative to single-focus scanning, but also achieves better localization in the heating pattern. The maximization of intensity gain of multiple-focus heating patterns significantly reduces the pre-focal-depth high-temperature regions that can be caused by single-focus scanning. This is shown by computer simulation of a two-dimensional cylindrical-section array (CSA2D) as a heating applicator. Two series of simulations are presented in which different scan trajectories were used to therapeutically heat a small deep-seated target volume. In every case the heating pattern was generated using single-focus scanning and multiple-focus scanning (with and without intensity gain maximization). Multiple-focus scanning with gain maximization offers the best localization of heating to the target volume of the three methods.
    [Abstract] [Full Text] [Related] [New Search]