These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Capsule endoscopy and cardiac pacemakers: investigation for possible interference. Author: Bandorski D, Irnich W, Brück M, Beyer N, Kramer W, Jakobs R. Journal: Endoscopy; 2008 Jan; 40(1):36-9. PubMed ID: 18067067. Abstract: BACKGROUND AND STUDY AIMS: The diagnostic benefit of capsule endoscopy in suspected small-bowel disease is high and seems to be superior to that with conventional methods and scintigraphy. Because of the limited clinical experience in patients with cardiac pacemakers and other implanted electrical devices, it is recommended that capsule endoscopy should not be used in such cases. We investigated the potential for capsule endoscopy to interfere with cardiac pacemakers in vitro. MATERIAL AND METHODS: 21 different pacemakers and leads were positioned in a 0.9 % saline solution in a tank. Pin jacks were placed that were in contact with the solution. The pacemaker impulse was recorded and an inhibiting signal could be coupled in. The capsules (Given Imaging and Olympus) were placed in different positions relative to the cardiac pacemaker and finally placed on the case of the pacemaker, for 1 minute in each site. The functioning of the pacemaker was observed continuously. The effect on the pacemaker was noted particularly as inhibition, synchronization, or no effect. The pacemaker was then inhibited using a standard test signal. RESULTS: There was no interference between the video capsule and the cardiac pacemakers in our experiment in spite of the close proximity of the two devices. CONCLUSION: The clinical use of capsule endoscopy is unproblematic in patients with cardiac pacemakers.[Abstract] [Full Text] [Related] [New Search]