These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Sensitive determination of ranitidine in rabbit plasma by HPLC with fluorescence detection.
    Author: Khedr A.
    Journal: J Chromatogr B Analyt Technol Biomed Life Sci; 2008 Feb 01; 862(1-2):175-80. PubMed ID: 18166505.
    Abstract:
    A sensitive high-performance liquid chromatographic method for determination of ranitidine (RAN) in rabbit plasma is described. The method is based on liquid-liquid extraction, labeling with dansyl chloride and monitoring with fluorescence detector at 338nm (ex)/523nm (em). Plasma samples were extracted with diethyl ether alkalinized with 1M sodium hydroxide. Ephedrine HCl (EPH-HCl) was used as internal standard. Both, RAN and EPH were completely derivatized after heating at 60 degrees C for 10min in sodium bicarbonate solution (pH 9.5). The derivatized samples were analyzed by HPLC using Agilent Zorbax Extended C18 column (150mmx4.6mm i.d.) and mobile phase consists of 48% acetonitrile and 52% sodium acetate solution (0.02M, pH 4.6). The linearity of the method was in the range of 0.025-10microg/ml. The limits of detection (LOD) and quantification (LOQ) were 7.5+/-0.18 and 22.5+/-0.12ng/ml, respectively. Ranitidine recovery was 97.5+/-1.1% (n=6; R.S.D.=1.8%). The method was applied on plasma collected from rabbits at different time intervals after oral administration of 5mg/kg ranitidine HCl.
    [Abstract] [Full Text] [Related] [New Search]