  • Title: Tone-evoked ABR in full-term and preterm neonates with normal hearing.
    Author: Ribeiro FM, Carvallo RM.
    Journal: Int J Audiol; 2008 Jan; 47(1):21-9. PubMed ID: 18196483.
    Abstract:
    This study was designed to investigate the feasibility of applying tone-ABRs in the nursery and neonatal intensive care unit (NICU), and to provide normative tone-ABR data from neonates. Normative tone-ABR latency data were determined. The study obtained intensity series of tone-ABRs from thirty preterm neonates and twenty full-term neonates who had confirmed normal peripheral auditory function after passing both an OAE and ABR screening examination. ABRs were collected in response to 500, 1500, and 4000 Hz tone bursts at 70, 50, 30, and 20 dB nHL. Mean wave V latencies were compared between groups, ears, and by gender. Responses to tone bursts of 20 and 30 dB nHL were detected in 97% and 100% of all ears respectively, in addition to responses to the higher-intensity stimuli. Preterm neonates' ABRs showed significantly longer latencies than those of the full-term infants. Tone-ABR evaluation was found to be both feasible and reliable as a measure of auditory function in neonates.