These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Experimental investigation of finite amplitude distortion-based, second harmonic pulse echo ultrasonic imaging.
    Author: Christopher T.
    Journal: IEEE Trans Ultrason Ferroelectr Freq Control; 1998; 45(1):158-62. PubMed ID: 18244167.
    Abstract:
    The computational predictions for the imaging potential of the second harmonic produced by finite amplitude distortion were investigated with a simple experiment. A focused transducer containing concentric 2.5 MHz and 5.0 MHz elements was used to obtain a sequence of radio-frequency (r-f) backscattered signals using a tissue equivalent phantom. The 2.5 MHz element was used as the transmitter and the 5.0 MHz element was used as the receiver. At 0.68 cm in front of the geometric focal point of the transducer, the phantom contained a 0.6 cm diameter cylindrical volume which contained no scatterers. Each of these r-f signals was then processed to produce the corresponding fundamental (2.5 MHz-centered) and second harmonic (5.0 MHz-centered) envelopes. The contrast resolution obtained for the scatterer-free or cyst region of the envelopes was compared against the computed prediction and good agreement was obtained. The results of this experiment also suggest that the simple one-pulse scheme may be adequate for second harmonic imaging.
    [Abstract] [Full Text] [Related] [New Search]