These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A new geometric factor for in situ resistivity measurement using four slender cylindrical electrodes. Author: Chong CE, Tan YL. Journal: IEEE Trans Biomed Eng; 2008 Feb; 55(2 Pt 1):594-602. PubMed ID: 18269995. Abstract: The four-electrode method is commonly used for in situ measurement of the electrical resistivity of biological tissues. In this paper, a new geometric factor between the resistivity and measured resistance using the four-electrode interface is derived in the prolate spheroidal coordinates and experimentally validated. Evaluation of the experimental results shows that the resistivities determined using both the derived geometric factor and a commercial conductivity meter are in close agreement even when the length of the immersed electrodes becomes long with respect to the inter-electrode spacing. The evaluation also shows the effect of the relative size of the sample volume when the limitation to semi-infinite volume begins to result in poor accuracy.[Abstract] [Full Text] [Related] [New Search]