Search MEDLINE/PubMed
Title: A comparison of two- and four-electrode techniques to characterize blood impedance for the frequency range of 100 Hz to 100 MHz. Author: Chang ZY, Pop GM, Meijer GM. Journal: IEEE Trans Biomed Eng; 2008 Mar; 55(3):1247-9. PubMed ID: 18334424. Abstract: Measurement setups that characterize the impedance of suspensions of blood over the wide frequency range of 100 Hz to 100 MHz are presented in this paper. The performance of the two- and four-electrode techniques have been compared and evaluated. By applying a combination of the two measurement techniques the best result is achieved when taking into account the main nonidealities, such as electrode polarization impedance and parasitic capacitances. It has been found that the conventional three-element model for the impedance of blood can be used for frequencies up to 1 MHz. For frequencies exceeding 1 MHz, an extended model is introduced where a constant phase angle element is used for modeling the cell membrane and a capacitor C(liq) is added for modeling the electrical capacitance of water in blood.[Abstract] [Full Text] [Related] [New Search]