Search MEDLINE/PubMed
Title: Determination of oxatomide in human plasma by high-performance liquid chromatography-electrospray ionization mass spectrometry. Author: She X, Wu L, Wei H, Liu W, Chen Y. Journal: Biomed Chromatogr; 2008 Jul; 22(7):746-52. PubMed ID: 18348338. Abstract: A rapid, sensitive and specific high-performance liquid chromatography-electrospray ionization mass spectrometry (LC-ESI-MS) method has been developed and validated for the determination of oxatomide in human plasma. Flunarizine hydrochloride was employed as the internal standard (IS). The analytes were chromatographically separated on a Shimadzu Shim-pack VP-ODS C18 column (250 x 2.0 mm i.d.) with a mobile phase consisting of methanol and aqueous ammonium acetate solution (10 mm, pH 4.0; 85:15, v/v). Detection was performed on a single quadrupole mass spectrometer using an electrospray ionization interface with the selected-ion monitoring (SIM) mode. The method showed excellent linearity (r = 0.9995) over the concentration range of 0.5-500 ng/mL with good accuracy and precision. The intra- and inter-batch precisions were within 10% relative standard deviation. The recoveries were more than 90%. The validated method was successfully applied to a preliminary pharmacokinetic study of oxatomide in Chinese healthy male volunteers.[Abstract] [Full Text] [Related] [New Search]