These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: alpha-Particle and gamma-ray spectrometry of a plutonium solution for impurity determination. Author: Sibbens G, Altzitzoglou T, Benedik L, Pommé S, Van Ammel R. Journal: Appl Radiat Isot; 2008; 66(6-7):813-8. PubMed ID: 18356064. Abstract: A highly enriched (240)Pu solution was measured by alpha-particle and gamma-ray spectrometry to determine other radionuclides present in the material as impurities. Low activities of (238)Pu, (241)Am, (243)Cm and (244)Cm were determined by measuring thin sources, made from the original solution, in a high-resolution alpha-particle spectrometer. The sources were prepared by evaporating the plutonium solution on quartz plates in a vacuum chamber. From the ingrowth of (241)Am in the original solution, the amount of (241)Pu could be calculated. After radiochemical separation of (241)Am, the plutonium was measured by high-efficiency alpha-particle spectrometry to determine the amount of (238)Pu. The enriched (240)Pu material was also measured by high-resolution gamma-ray spectrometry, using two different HPGe detectors to determine the impurities of (239)Pu and (241)Am. The preparation of the sources and the measurement methods are described and discussed. The measured impurities, given in % of the (240)Pu activity, are compared with the values on the certificate.[Abstract] [Full Text] [Related] [New Search]