These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Quantification of cyclosporine and tacrolimus in whole blood. Comparison of liquid chromatography-electrospray mass spectrometry with the enzyme multiplied immunoassay technique.
    Author: Ansermot N, Fathi M, Veuthey JL, Desmeules J, Rudaz S, Hochstrasser D.
    Journal: Clin Biochem; 2008 Jul; 41(10-11):910-3. PubMed ID: 18358833.
    Abstract:
    OBJECTIVES: The aim of this work was to compare a validated liquid chromatography-mass spectrometry (LC-MS) method with the commercial enzyme multiplied immunoassay technique (EMIT) for cyclosporine and tacrolimus whole blood quantification. DESIGN AND METHODS: Samples of transplant patients receiving cyclosporine (n=38) or tacrolimus (n=41) were analyzed successively by LC-MS and EMIT. Several statistical approaches for method comparison were evaluated and Passing-Bablok and Bland-Altman analyses chosen. RESULTS: Overestimations of the concentrations measured with EMIT compared to LC-MS were observed with means of 23% (range: 6% to 46%) for cyclosporine and 30% (range: -3% to 73%) for tacrolimus. CONCLUSION: The EMIT demonstrated significant positive biases due to cross-reactions with metabolites. This indicates that, in some clinical situations, a selective method such as LC-MS is preferable for therapeutic drug monitoring in transplant patients.
    [Abstract] [Full Text] [Related] [New Search]