These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Enzymatic probe sonication as a tool for solid-liquid extraction for total selenium determination by electrothermal-atomic absorption spectrometry.
    Author: Vale G, Pereira S, Mota A, Fonseca L, Capelo JL.
    Journal: Talanta; 2007 Nov 30; 74(2):198-205. PubMed ID: 18371630.
    Abstract:
    A new fast and reproducible approach is described for the application of the enzymatic probe sonication (EPS) methodology [J.L. Capelo, P. Ximénez-Embún, Y. Madrid-Albarrán, C. Cámara, Anal. Chem. 76 (2004) 233-237] for total selenium determination by electrothermal atomic absorption spectrometry, ET-AAS. Ni(NO3)2 and Pd(NO3)2 were studied as matrix modifiers in conjunction with H2O2, being best results obtained with Pd(NO3)2 plus H2O2. The presence of H2O2 as matrix modifier increases up to 66% the time-life of the graphite tubes, by avoiding the building-up of carbonaceous residues. BCR-414 plankton and ERM-CE 278 mussel tissue reference materials were used for proof-of-the-methodology. Different enzymes, protease XIV, substilisin and trypsin were studied. The use of fresh enzyme was found critical. Good Se recoveries were obtained for oyster tissue, 111%; BCR-414 plankton, 106% and ERM-CE 278 mussel tissue, 93%, when protease XIV was used. Data regarding microwave digestion versus EPS methodology is also presented and discussed.
    [Abstract] [Full Text] [Related] [New Search]