These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Severe tricuspid stenosis during percutaneous occlusion of perimembranous ventricular septal defect with the new Amplatzer device.
    Author: Christiani LA, Bergman F, Tress JC, Vanzillotta PP, Pedra CA.
    Journal: Congenit Heart Dis; 2006 Sep; 1(5):239-43. PubMed ID: 18377532.
    Abstract:
    BACKGROUND: The new Amplatzer device has been used with success for occlusion of the perimembranous ventricular septal defects and the index of complications is very low. RESULTS: The reported case corresponds to a very rare severe acute tricuspid stenosis, caused by adherence of the prosthesis to the anterior tricuspid leaflet. CONCLUSION: Device release was achieved by manipulation on the tricuspid apparatus.