These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Experimental measurements and Monte Carlo calculations of dosimetric parameters of the IRA1-103Pd brachytherapy source.
    Author: Sadeghi M, Hosseini SH, Raisali G.
    Journal: Appl Radiat Isot; 2008 Oct; 66(10):1431-7. PubMed ID: 18387806.
    Abstract:
    This work presents a brachytherapy source having (103)Pd adsorbed onto a cylindrical silver rod that has been developed by Agricultural, Medical and Industrial Research School for permanent implant applications. Dosimetric characteristics (dose-rate constant, radial dose function, anisotropy function and anisotropy factor) of this source were experimentally and theoretically determined in terms of the updated AAPM Task Group 43 (TG-43U1) recommendations. Measurements were performed using TLD-GR200A circular chip dosimeters using standard methods employing thermoluminescent dosimeters in a Perspex phantom. Precision machined bores in the phantom located dosimeters and source in a reproducible fixed geometry providing for transverse-axis and angular dose profiles over a range of distances from 0.5 to 5 cm. The Monte Carlo N-Particle (MCNP) code, version 4C was used to evaluate the dose-rate distributions around this model (103)Pd source in water and Perspex phantoms. The Monte Carlo calculated dose-rate constant of the IRA1-(103)Pd source in water was found equal to Lambda=0.669 cGy/h/U with approximate uncertainties of +/-0.1%. The anisotropy function, F(r, theta), and the radial dose function, g(L)(r), of the IRA1-(103)Pd source were also measured in Perspex phantom and calculated in both Perspex and liquid water phantom.
    [Abstract] [Full Text] [Related] [New Search]