These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Perchlorate-selective polymeric membrane electrode based on bis(dibenzoylmethanato)cobalt(II) complex as a neutral carrier. Author: Rezaei B, Meghdadi S, Bagherpour S. Journal: J Hazard Mater; 2009 Jan 30; 161(2-3):641-8. PubMed ID: 18486336. Abstract: A synthesized bis(dibenzoylmethanato)Co(II) complex (Co(DBM)(2)), has been used as a ionophore for the preparation of a new perchlorate ion-selective electrode. The electrode exhibits a Nernstian response over the perchlorate concentration range of 8.0x10(-7)-1.0x10(-1)M with a slope of 60.3+/-0.5 mV per decade of concentration. The limit of detection as determined from the intersection of the extrapolated linear segments of the calibration plot is 5.6x10(-7)M. The electrode shows good selectivity towards perchlorate with respect to many common anions. The response time of the sensor is very fast (< or = 5s), and can be used for at least 2 months in the pH range of 2.0-9.0. The electrode was used to determine perchlorate in water and human urine. The interaction of the ionophore with perchlorate ions was demonstrated by UV-vis spectroscopy.[Abstract] [Full Text] [Related] [New Search]