These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Prostate MR imaging at 3T with a longitudinal array endorectal surface coil and phased array body coil.
    Author: Kim DY, Schnall MD, Rosen MA, Connick T.
    Journal: J Magn Reson Imaging; 2008 Jun; 27(6):1327-30. PubMed ID: 18504745.
    Abstract:
    PURPOSE: To demonstrate the feasibility of using a double loop phased array endorectal coil combined with a phased array body coil to image the prostate at 3T. MATERIALS AND METHODS: We designed and constructed a novel prostate coil employing two arrayed 4.0 x 5.0 cm loops, tuned the device for optimal performance at 3T, and characterized the signal-to-noise ratio (SNR) associated with it. RESULTS: The coil Q factor was calculated to be approximately 50 unloaded, and 30 when loaded on human tissue. SNR maps at multiple orientations were constructed and images were acquired on both a phantom and a human. As expected, SNR was highest along the midpoint of the array and demonstrated strong signal even at 4 cm from the coil. CONCLUSION: The double loop phased array endorectal coil combined with a phased array body coil at 3T is feasible in vivo and compelling enough to warrant future clinical trials to evaluate its efficacy. These trials are currently under way.
    [Abstract] [Full Text] [Related] [New Search]