Title: Prostate MR imaging at 3T with a longitudinal array endorectal surface coil and phased array body coil. Author: Kim DY, Schnall MD, Rosen MA, Connick T. Journal: J Magn Reson Imaging; 2008 Jun; 27(6):1327-30. PubMed ID: 18504745. Abstract: PURPOSE: To demonstrate the feasibility of using a double loop phased array endorectal coil combined with a phased array body coil to image the prostate at 3T. MATERIALS AND METHODS: We designed and constructed a novel prostate coil employing two arrayed 4.0 x 5.0 cm loops, tuned the device for optimal performance at 3T, and characterized the signal-to-noise ratio (SNR) associated with it. RESULTS: The coil Q factor was calculated to be approximately 50 unloaded, and 30 when loaded on human tissue. SNR maps at multiple orientations were constructed and images were acquired on both a phantom and a human. As expected, SNR was highest along the midpoint of the array and demonstrated strong signal even at 4 cm from the coil. CONCLUSION: The double loop phased array endorectal coil combined with a phased array body coil at 3T is feasible in vivo and compelling enough to warrant future clinical trials to evaluate its efficacy. These trials are currently under way.[Abstract] [Full Text] [Related] [New Search]