These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: In-vitro testing of three totally supra-annular bileaflet mechanical valves: hydrodynamics in the Sheffield pulse duplicator.
    Author: Bottio T, Tarzia V, Rizzoli G, Gerosa G.
    Journal: J Heart Valve Dis; 2008 Mar; 17(2):222-6. PubMed ID: 18512495.
    Abstract:
    BACKGROUND AND AIM OF THE STUDY: In-vivo echocardiographic studies are limited by several confounding factors and technical pitfalls, and consequently the hypothetical differential hydraulic behavior between different prosthetic heart valves has not been identified. However, for surgeons it is essential to know the functional and geometric characteristics of the prostheses to be used. Herein, the in-vitro performance of two new supra-annular bileaflet prostheses--the Medtronic Advantage Supra and Sorin Bicarbon Overline--was compared with that of the 21-mm St. Jude Medical (SJM) Regent valve. METHODS: Three high-performance, production-quality prostheses, including the sewing-ring cuffs, were tested in the aortic chamber of a Sheffield pulse duplicator. The sizes of the prostheses which fitted the 21-mm valve holder were: 21 mm Advantage Supra Medtronic; 19 mm SJM Regent, and 18 mm Sorin Bicarbon Overline. The tests were carried out at a fixed pulse rate (70 beats/min), and at increasing cardiac output (CO) of 2, 4, 5, and 7 l/min. Forward-flow pressure drop, total regurgitant volume, closing and leakage volumes, effective orifice area (EOA) and stroke work loss (SWL) were recorded while the valve was operated at each CO. RESULTS: The SJM Regent and Sorin Bicarbon Overline valves each showed, at increasing CO, significantly lower mean and peak gradients. The calculated EOA and SWL were significantly better with the SJM Regent and Sorin Bicarbon Overline prostheses. The Medtronic Advantage Supra valve showed comparable results only while performing at 2 l/min CO. With regards to the regurgitant fraction, lowest values were observed with the Medtronic Advantage Supra valve. CONCLUSION: This hydrodynamic evaluation model allowed a comparison to be made of the efficiency of recently commercialized bileaflet prostheses, among which the older SJM Regent and the newer Sorin Bicarbon Overline valves demonstrated the best performances.
    [Abstract] [Full Text] [Related] [New Search]