These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Picometer positioning system based on a zooming interferometer using a femtosecond optical comb.
    Author: Kajima M, Matsumoto H.
    Journal: Opt Express; 2008 Feb 04; 16(3):1497-506. PubMed ID: 18542225.
    Abstract:
    A positioning system based on a new zooming interferometer was developed for calibrating ultrahigh-resolution linear encoders. In the system, the positioning resolution is scaled down from that of the driving stage to that of the controlled stage by a zooming ratio that is determined by the ratio between the wavelengths of the optical sources for the zooming interferometer, so that the resolution in the controlled stage becomes very high. A femtosecond optical comb is used to stabilize two external-cavity diode lasers that are used as optical sources for the interferometer. The system realized a resolution of better than 30 pm and a stability of 1 nm.
    [Abstract] [Full Text] [Related] [New Search]