These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Stress distribution on external hexagon implant system using 3d finite element analysis.
    Author: Segundo RM, Oshima HM, Silva IN, Júnior LH, Mota EG, Coelho LF.
    Journal: Acta Odontol Latinoam; 2007; 20(2):79-81. PubMed ID: 18590255.
    Abstract:
    The aim of this study was to compare and evaluate strain distribution on dental implant, abutment, screw and crown virtual models in the posterior region. The analysis was performed by means of a 3D virtual model developed by the PRO-ENGINEER System (PRO-ENGINEER, PTC, Needham, MA, USA ) with an external butt joint (3i Implant Innovations, Palm Beach, Florida), square headed Gold Tite abutment retainer screw (3i Implant Innovations, Palm Beach, Florida), STA abutment (3i Implant Innovations, Palm Beach, Florida), metal infrastructure of Ag-Pd alloy and feldspatic ceramic. The standard load was 382N at 15 degree angle to the implant axis, applied at 6 mm from the implant center at different observation points on the implant-screw set. The data showed that on the implant virtual model, the highest strain concentration was found at the interface between the implant platform and the abutment, and in the middle point of the 1st screw thread internal diameter on the load application side.
    [Abstract] [Full Text] [Related] [New Search]