    Abstract:
    The Sysmex XS-1000i is a compact new, fully automated haematology analyser, designed to generate complete blood counts with five-part leucocyte differential. In our study, a Sysmex XS-1000i instrument was evaluated according to Clinical Laboratory Standards Institute (CLSI) and International Council for Standardization in Haematology (ICSH) guidelines. Precision, carry-over and linearity were determined. Using a total of 700 patient samples, results from the Sysmex XS-1000i were compared with those from a Sysmex XE-2100, an Abbott Cell Dyn 4000 and the manual reference leucocyte differential. Using quality control material, total and within-run imprecision was less than 3% except for platelets. The system demonstrated good linearity over the entire reporting range and no carry-over (<0.5%). The Sysmex XS-1000i showed good correlation with XE-2100, CD-4000 and the manual reference leucocyte differential. Overall flagging sensitivity and specificity were 91% and 48%, respectively. In conclusion, the Sysmex XS-1000i demonstrated good analytical performance, is able to generate a complete blood count with five-part differential on low blood volumes and has considerable back-up capacity.