These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A high peak current 150-J fixed-energy defibrillation protocol treats recurrent ventricular fibrillation (VF) as effectively as initial VF. Author: Hess EP, Russell JK, Liu PY, White RD. Journal: Resuscitation; 2008 Oct; 79(1):28-33. PubMed ID: 18621462. Abstract: OBJECTIVE: We tested the hypothesis that the frequency of shock success differs between initial and recurrent episodes of ventricular fibrillation (VF). METHODS: Out-of-hospital cardiac arrest patients presenting with VF from December 1996 to February 2007 defibrillated using an AED with a fixed-energy protocol (150 J) (Philips Medical Systems, Seattle, WA) were included for analysis. We defined shock success as termination of VF within 5s post-shock (either asystole or organized rhythm). Generalized estimating equation (GEE) analysis was used to adjust for the interrelatedness of shocks within each patient. RESULTS: One hundred and three events occurred during the study period. Patient characteristics included: mean age 64.4 years, 82.5% male, and 81.6% bystander witnessed. Synchronized call-to-shock time was 6.4+/-2.3 min (mean+/-S.D.). VF recurred in 64 (62.1%) patients. Two hundred and fifty-seven shocks delivered for initial (101) or recurrent (156) VF were available for analysis. Initial shocks terminated VF in 93/101 (92.1%); subsequent shocks terminated recurrent VF in 140/156 (89.7%). GEE odds ratio for shock type (initial versus refibrillation) was 1.10 (95% CI 0.37-3.24, p=0.87). After adjusting for potential confounders, shock type remained insignificant (OR 1.14, 95% CI 0.41-3.2, p=0.80). We observed no significant difference in ROSC (34.4% versus 46.2%, p=0.23) or survival (37.5% versus 41.0%, p=0.72) between those with and without VF recurrence. CONCLUSIONS: We observed no significant difference in the frequency of shock success between initial and recurrent episodes of VF using this AED with a 150 J fixed-energy protocol. VF recurrence is common and does not adversely affect shock success, ROSC or survival.[Abstract] [Full Text] [Related] [New Search]