These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Enhanced sensitivity for Cu(II) by a salicylidine-functionalized polysiloxane carbon paste electrode.
    Author: Abu-Shawish HM, Saadeh SM, Hussien AR.
    Journal: Talanta; 2008 Aug 15; 76(4):941-8. PubMed ID: 18656682.
    Abstract:
    A new approach for decreasing the detection limit for a copper(II) ion-selective electrode (ISE) is presented. The ISE is designed using salicylidine-functionalized polysiloxane in carbon paste. This work describes the attempts to develop the electrode and measurements of its characteristics. The new type of renewable three-dimensional chemically modified electrode could be used in a pH range of 2.3-5.4, and its detection limit is 2.7 x 10(-8)mol L(-1) (1.2 microg L(-1)). This sensor exhibits a good Nernstian slope of 29.4+/-0.5 mV/decade in a wide linear concentration range of 2.3 x 10(-7) to 1.0 x 10(-3)mol L(-1) of Cu(II). It has a short response time (approximately 8s) and noticeably high selectivity over other Cu(II) selective electrodes. Finally, it was satisfactorily used as an indicator electrode in complexometric titration with EDTA and determination of copper(II) in miscellaneous samples such as urine and various water samples.
    [Abstract] [Full Text] [Related] [New Search]