These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of betamethasone in human plasma by liquid chromatography with tandem mass.
    Author: Qu TT, Zhang R, Wang BJ, Liu XY, Yuan GY, Guo RC.
    Journal: Yao Xue Xue Bao; 2008 Apr; 43(4):402-7. PubMed ID: 18664204.
    Abstract:
    A sensitive and selective liquid chromatography-tandem mass spectrometric (LC-MS/MS) method was developed and validated for the determination of betamethasone in human plasma. The analyte was isocratically eluted on a Venusil XBP C8 column (200 mm x 3.9 mm ID, 5 microm) with methanol-water mol x L(-1) ammonium formate) (80:20) at a flow rate of 0.4 mL x min(-1), and detected (containing 5 mmol x L(-1) ammonium formate) (80:20) at a flow rate of 0.4 mL x min(-1), and detected with a triple quad LC-MS/MS using ESI with positive ionization. Ions monitored in the multiple reaction monitoring (MRM) mode were m/z 393.3-->355.2 for betamethasone and m/z 361.3-->343.2 for prednisolone (IS). Betamethasone was extracted from 0.5 mL human plasma with ethyl acetate. The average recovery is 88.24% and the low limit of quantitation (LLOQ) was 0.5 ng x mL(-1). The 3-day validation study demonstrated excellent precision and accuracy across the calibration range of 0.5-80.0 ng x mL(-1). The method was successfully applied to the pharmacokinetic study of compound betamethason injection in healthy Chinese volunteers.
    [Abstract] [Full Text] [Related] [New Search]