These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Performance characteristics of the Access Inhibin A assay.
    Author: Rawlins ML, La'ulu SL, Erickson JA, Roberts WL.
    Journal: Clin Chim Acta; 2008 Nov; 397(1-2):32-5. PubMed ID: 18691567.
    Abstract:
    BACKGROUND: Second trimester maternal screening using AFP, uE3, hCG, and inhibin A has shown a detection rate for Down's syndrome of 81% with a 5% false positive rate. Inhibin A may also have utility as a serum tumor marker in postmenopausal women with ovarian cancer and men with testicular stromal tumors. METHODS: The Beckman Coulter Access Inhibin A assay was evaluated for limit of blank, dilution linearity, imprecision, interferences, reference intervals, and comparison to an inhibin A ELISA. RESULTS: The limit of blank was 0.1 ng/l. The assay was linear from 0.2 to 1347 ng/l. Total inter-assay CVs were <5% for control levels ranging from 24.6 ng/l to 811 ng/l. Interference studies showed recoveries of inhibin A within 10% of expected values at interferent concentrations of 10 g/l hemoglobin and 22 g/l triglycerides. No significant interference was observed at a bilirubin concentration of 400 mg/l. The 97.5th percentile upper reference limits were 6.8 ng/l for postmenopausal women and 3.0 ng/l for men. The Access assay compared to an ACTIVE ELISA showed a slope of 0.88, an intercept of -3.7 ng/l, S(y/x)=40 ng/l, and r=0.98. CONCLUSIONS: The analytical performance of the Access inhibin A assay is acceptable for routine laboratory testing.
    [Abstract] [Full Text] [Related] [New Search]