These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [QRS detection based on difference method].
    Author: Chen D, Zhang B, Yang S.
    Journal: Sheng Wu Yi Xue Gong Cheng Xue Za Zhi; 2008 Jun; 25(3):502-5. PubMed ID: 18693419.
    Abstract:
    A difference method for detection of QRS is proposed in this paper. By calculating the difference function of the Electrocardiogram (ECG), the method can eliminate or weaken the impact of the P wave, T wave and other interference signals onto the detection of QRS. It can solve the problem of low detection accuracy yielded by the traditional method, and can solve the problem of heavily loaded computation caused by the complicated algorithm of the wavelet method. Experiments using data from the MIT-BIH arrhythmia database illustrate that this method has good adaptability for various ill-conditioned ECG waves. This method has found clinical applications in the '12 lead ECG system' developed by the authors. The detection accuracy rate is higher than 99%.
    [Abstract] [Full Text] [Related] [New Search]