These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Digital noise reduction: outcomes from laboratory and field studies.
    Author: Bentler R, Wu YH, Kettel J, Hurtig R.
    Journal: Int J Audiol; 2008 Aug; 47(8):447-60. PubMed ID: 18698521.
    Abstract:
    The purpose of this study was to determine the impact of a digital noise reduction (DNR) scheme implemented in a current commercial hearing aid. In a double-blinded design, three conditions of onset time (4, 8, 16 seconds) were randomly assigned to the 25 subjects, plus one condition wherein the noise-reduction feature was disengaged. Subsequently, a fifth trial/condition, wherein the subject had access to three memories in which the different onsets were programmed, was carried out. For each of the five conditions, the subjects had an at-home trial, prior to obtaining self-report measures. Laboratory measures of speech perception showed no effect of the DNR, with or without the provision of visual cues. Laboratory-based ratings of ease of listening showed DNR-on (all onset times) to be rated significantly better than DNR-off; for ratings of listening comfort, the 4-second onset time was rated significantly lower (poorer) than the 8-second onset or the DNR-off condition; for ratings of sound quality, DNR-on or -off had no differential effect. Self-report measures indicated significantly higher aversiveness in the DNR-off condition compared to the pre-test scores.
    [Abstract] [Full Text] [Related] [New Search]