These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Motion artifact simulating aortic dissection on CT.
    Author: Burns MA, Molina PL, Gutierrez FR, Sagel SS.
    Journal: AJR Am J Roentgenol; 1991 Sep; 157(3):465-7. PubMed ID: 1872227.
    Abstract:
    We recently imaged two patients clinically suspected of having aortic dissection whose contrast-enhanced CT examinations, obtained on a new scanner with a 1-sec scanning time, showed findings suggesting an ascending aortic dissection. The subsequent clinical course and evaluation implied that the CT findings were predominantly artifactual. We identified identical artifacts in 18% of 50 consecutive contrast-enhanced CT examinations performed for a variety of indications on the same scanner. The double-lumen artifact, simulating an intimal flap, occurs in the proximal ascending aorta and is limited to one or two contiguous transaxial images. The artifact was not detected on two other CT units. We believe the artifact arises from motion of the aortic wall and the surrounding pericardial recesses during image acquisition.
    [Abstract] [Full Text] [Related] [New Search]