These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Diffusion tensor imaging and fiber tractography of the median nerve at 1.5T: optimization of b value. Author: Andreisek G, White LM, Kassner A, Tomlinson G, Sussman MS. Journal: Skeletal Radiol; 2009 Jan; 38(1):51-9. PubMed ID: 18773203. Abstract: OBJECTIVE: The objective of this study was to systematically assess the optimal b value for diffusion tensor imaging and fiber tractography of the median nerve at 1.5 T. MATERIALS AND METHODS: This is a prospective study which was carried out with institutional review board approval and written informed consent from the study subjects. Fifteen healthy volunteers (seven men, eight women; mean age, 31.2 years) underwent diffusion tensor imaging of the wrist. A single-shot spin-echo-based echo-planar imaging sequence (TR/TE, 7000/103 ms) was performed in each subject at eight different b values ranging from 325 to 1,550 s/mm(2). Number and length of reconstructed fiber tracts, fiber density index (FDi), fractional anisotropy (FA), and apparent diffusion coefficient (ADC) were calculated for the median nerve. Signal-to-noise ratio (SNR) was also calculated for each acquisition. The overall image quality was assessed by two readers in consensus by ranking representative fiber tract images for each subject using a scale range from 1 to 8 (1 = best to 8 = worst image quality). RESULTS: Longest fibers were observed for b values between 675 and 1,025 s/mm(2). Maximum FDi was found at b values of 1,025 s/mm(2). FA was between 0.5 and 0.6 for all b values. ADC gradually decreased from 1.44 x 10(-3) to 0.92 x 10(-3) mm(2)/s with increasing b values. Maximum SNR +/- standard deviation (175.4 +/- 72.6) was observed at the lowest b value and decreased with increasing b values. SNR at b values of 1,025 s/mm(2) was 48.5% of the maximum SNR. Optimal fiber tract image quality was found for b values of 1,025 s/mm(2). CONCLUSIONS: The optimal b value for diffusion tensor imaging and fiber tractography of the median nerve at 1.5 T was 1,025 s/mm(2).[Abstract] [Full Text] [Related] [New Search]