These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Basic evaluation of an ambulatory ventricular function monitor with a a cadmium telluride detector].
    Author: Taki J, Muramori A, Nakajima K, Bunko H, Matsunari I, Tonami N, Hisada K.
    Journal: Kaku Igaku; 1991 Apr; 28(4):371-9. PubMed ID: 1880971.
    Abstract:
    The reliability of a new ambulatory ventricular function monitor (VEST) with cadmium telluride (Cd-Te) detector was validated. No counting loss was observed under 20 kcps. In phantom study, more than 200 ml with 6 cm distance from the detector and more than 300 ml with 6-12 cm distance, underestimation of volume was observed. Ejection fraction (EF) calculated in phantom study showed close value to the true EF in 50-400 ml of end-diastolic volume. In 10 patients, 1-2 cm detector shift toward right upper, upper and left showed underestimation of EF. Good correlation of gamma camera EF (X) and VEST EF (Y) was obtained (Y = 0.96X + 3.6(%), n = 37, r = 0.93, p less than 0.001). EF change from rest values, obtained from each comparable exercise and post exercise stage by gamma camera (X) and VEST (Y), showed good correlation (Y = 0.95X + 0.8, n = 20, r = 0.85, p less than 0.001). In conclusion Cd-Te VEST was considered as a reliable EF measurement technique similar to gamma camera.
    [Abstract] [Full Text] [Related] [New Search]