These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Use of calcium-titrated sodium heparinate for ionized calcium determination on plasma and whole blood (heparinized blood sampler B-129 Radiometer).
    Author: Sachs C, Rabouine P, Chaneac M, Kindermans C, Dechaux M.
    Journal: Scand J Clin Lab Invest; 1991 May; 51(3):303-6. PubMed ID: 1882182.
    Abstract:
    We report modifications of ionized calcium as observed on specimens collected in a commercially available sampling device (Radiometer B-129). When the sampler is used as recommended, errors are non-significant around 1.25 mmol l-1. There is a +3% increase for low (0.75 mmol l-1) values and a -2% decrease for high (2.50 mmol l-1) values. Incomplete syringe-filling increases these errors.
    [Abstract] [Full Text] [Related] [New Search]