These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The natural history of metallosis from catastrophic failure of a polyethylene liner in a total hip.
    Author: Khan RJ, Wimhurst J, Foroughi S, Toms A.
    Journal: J Arthroplasty; 2009 Oct; 24(7):1144.e1-4. PubMed ID: 18848419.
    Abstract:
    We report on a case of metallosis initially presumed to be heterotopic ossification based on radiologic findings. A 68-year-old man with a total hip arthroplasty experienced failure of the polyethylene liner, resulting in articulation of the ceramic head with the titanium acetabular shell. During revision surgery, extensive metallic debris was evident macroscopically throughout the periprosthetic tissue and was confirmed histologically to be metallosis.