These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A highly sensitive spectrophotometric method for determination of micro amounts of arsenic. Author: Qian-Feng W, Peng-Fei L. Journal: Talanta; 1983 Apr; 30(4):275-6. PubMed ID: 18963357. Abstract: A highly sensitive spectrophotometric method for determination of arsenic, based on the formation of an ion-association complex between arsenoantimonomolybdenum blue and Malachite Green, has been developed. The ion-association complex is soluble in the presence of Triton X-305, so arsenic can be determined directly in aqueous solution. The apparent molar absorptivity for arsenic is 1.13 x 10(5) l.mole(-1).cm(-1) at 640 nm. Beer's law is obeyed for 0-5 mug of arsenic. The lower limit of determination (absorbance = 0.01) is 4 ng ml in the final solution.[Abstract] [Full Text] [Related] [New Search]