  • Title: Determination of ketorolac in human serum by square wave adsorptive stripping voltammetry.
    Author: Radi A, Beltagi AM, Ghoneim MM.
    Journal: Talanta; 2001 Apr 12; 54(2):283-9. PubMed ID: 18968250.
    Abstract:
    The adsorption behavior of ketorolac on a hanging mercury drop electrode (HMDE) was explored by square-wave and cyclic voltammetry. The square wave voltammetric response of ketorolac depends on the parameters of the square wave voltammetry excitation signal as well as on the pH of the medium and the accumulation time. The drug was accumulated at HMDE and a well-defined peak was obtained at -1.41 V versus. Ag/AgCl (saturated KCl) in acetate buffer of pH 5.0. A square-wave adsorptive stripping voltammetric method for the quantitative determination of ketorolac was developed. The linear concentration range was 1x10(-10)-1x10(-8) when using 300 s accumulation at -0.8 V. The detection limit of ketorolac was 1.0x10 (-11)M . The precision was excellent with relative standard deviation of 3.85% at concentration of 5x10 (-8)M after 60 s accumulation time. Applicability to serum samples was illustrated. A detection limit of 14 ng per ml of serum was obtained.