These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Use of Nb(2)O(5)-SiO(2) in an automated on-line preconcentration system for determination of copper and cadmium by FAAS.
    Author: da Silva EL, Ganzarolli EM, Carasek E.
    Journal: Talanta; 2004 Mar 10; 62(4):727-33. PubMed ID: 18969355.
    Abstract:
    A procedure for the determination of trace level of copper(II) and cadmium(II) by FAAS using an on-line preconcentration system has been proposed. In this system, copper and cadmium ions were adsorbed onto a minicolumn packed with silica gel modified with niobium(V) oxide (Nb(2)O(5)-SiO(2)), followed by nitric acid elution in reverse mode and determination on-line by flame atomic absorption spectrometry (AAS) without interference of the matrix. Chemical and flow variables as well as concomitant ions were studied in the developed procedure. The enrichment factor for copper(II) and cadmium(II) was 34.2 and 33.0, respectively, using a preconcentration time of 2min. The limit of detection for copper(II) and cadmium(II) was 0.4, and 0.1mugl(-1), respectively. The precision of the method, evaluated as the relative standard deviation in solutions containing 15mugl(-1) of copper and 10mugl(-1) of cadmium, by analyzing a series of seven replicates, was 1.8 and 1.6%, respectively. The accuracy was assessed through recovery experiments of certified material and water samples.
    [Abstract] [Full Text] [Related] [New Search]