These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Differential phase decoder in a polarized optical heterodyne interferometer.
    Author: Chou C, Teng HK, Tsai CC, Wu JS.
    Journal: J Opt Soc Am A Opt Image Sci Vis; 2008 Nov; 25(11):2630-5. PubMed ID: 18978839.
    Abstract:
    A differential-phase decoder (DPD) together with a polarization common-path optical heterodyne interferometer is set up. Based on this interferometric configuration and a novel balanced-detector scheme, the performance of the quantum-noise-limited differential-phase decoder is demonstrated and analyzed. The minimum-detectable differential phase is on the order of 10(-7) rad/sqrt Hz when a 2.5 mW He-Ne laser is used. Verified experimentally, the DPD is immune to the common-phase noise induced by an electro-optic phase modulator or by thermal disturbance within the interferometer. This signifies that the minimum-detectable differential phase can become 10(-8) rad/sqrt Hz if a 300 mW continuous wave laser is employed instead.
    [Abstract] [Full Text] [Related] [New Search]