These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Analysis and evaluation of a novel quality assurance device for ultrasonic medical imaging systems. Author: Hah Z, Naum R. Journal: IEEE Trans Ultrason Ferroelectr Freq Control; 2008 Oct; 55(10):2345-52. PubMed ID: 18986883. Abstract: A novel test device that has the potential to facilitate day-to-day maintenance and quality assurance of ultrasonic medical imaging systems was developed and tested. The device consists of small echogenic particles deposited on a thin film substrate in patterns similar to those found on conventional eye charts, which can be scanned to produce recognizable images that can be quantitatively evaluated in a quick and repeatable fashion. An evaluation system consisting of this device and a related assessment methodology was tested at several locations by both physicians and sonographers. The results of the testing and responses of the users were both encouraging and insightful.[Abstract] [Full Text] [Related] [New Search]