These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [RovoIving ultrasonic probe for medical endoscope imaging system based on USB2.0].
    Author: Chen X, Wen S, Yu D.
    Journal: Sheng Wu Yi Xue Gong Cheng Xue Za Zhi; 2008 Oct; 25(5):1048-52. PubMed ID: 19024444.
    Abstract:
    This paper mainly introduces the design and implementation of rotary scanning imaging for Endoscopic Ultrasonography System based on USB2.0. The ultrasonic pulse-echo imaging technique has been used. In the paper, we simply introduce the emission circuit, receiving circuit and isolation circuit. According to the character of rotary scanning, we design the synchronous control circuit based on FPGA and data transmission circuit based on USB2.0. Finally the original polar coordinate image is transformed to rectangular coordinate grey image through coordinate transformation. The system performances have been validated by the experimental result gotten by scanning a specific object with continuous rotary motor.