These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of intact oxaliplatin in human plasma using high performance liquid chromatography-tandem mass spectrometry.
    Author: Zhang W, Seymour L, Chen EX.
    Journal: J Chromatogr B Analyt Technol Biomed Life Sci; 2008 Dec 15; 876(2):277-82. PubMed ID: 19026601.
    Abstract:
    A HPLC-tandem mass spectrometry method was developed and validated for the quantitation of intact oxaliplatin in human plasma. Plasma ultrafiltrates were precipitated with acetonitrile and separation was performed on a 250 mm Beckman ODS reverse phase column using a gradient mobile phase. The mass spectrometer was operated in positive ionization mode using TurboionSpray and precursor-product ion combinations of m/z 391.1-->305.1 and 371.1-->247.0 were monitored for oxaliplatin and carboplatin, the internal standard, respectively. The lower limit of quantitation for oxaliplatin was 20 ng/ml. The linear range of the method was 20-1000 ng/ml. The between- and within-day relative standard deviations ranged from 3.1 to 7.7%, and accuracy was within 5%. This method was successfully applied in a clinical study of oxaliplatin.
    [Abstract] [Full Text] [Related] [New Search]