These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Factors leading to tracheobronchial self-expandable metallic stent fracture.
    Author: Chung FT, Lin SM, Chen HC, Chou CL, Yu CT, Liu CY, Wang CH, Lin HC, Huang CD, Kuo HP.
    Journal: J Thorac Cardiovasc Surg; 2008 Nov; 136(5):1328-35. PubMed ID: 19026824.
    Abstract:
    OBJECTIVE: This retrospective study was to determine factors that contribute to self-expandable metallic stent fracture in patients with tracheobronchial disease. METHODS: From 2001 to 2006, 139 patients (age, 62.1 +/- 15.4 years; range, 23-87 years) with benign (n = 62) and malignant (n = 77) tracheobronchial disease received 192 Ultraflex (Boston Scientific, Natick, Mass) self-expandable metallic stents (98 in patients with benign disease and 94 in patients with malignant disease). RESULTS: Seventeen fractured self-expandable metallic stents were found; the incidence was 12.2% (17/139 patients) among patients with tracheobronchial disease. Tortuous airway (odds ratio, 4.06; 95% confidence interval, 1.04-18.34; P = .04) independently predicted self-expandable metallic stent fracture. Most self-expandable metallic stent fractures (64.7%, 11/17) were detected 500 to 1000 days after self-expandable metallic stent implantation. Clinical presentations for patients with fractured self-expandable metallic stents included dyspnea exacerbation (70.6%, 12/17) and cough (23.5%, 4/17). CONCLUSIONS: Self-expandable metallic stent fracture is not uncommon in patients with tracheobronchial disease. Tortuous airway is an independent predictor for it. Although management of the fractured self-expandable metallic stent in our study was feasible and safe, self-expandable metallic stents should be restricted to a more select population.
    [Abstract] [Full Text] [Related] [New Search]