These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Extravascular lung water measurement using transpulmonary thermodilution in children.
    Author: Lemson J, Backx AP, van Oort AM, Bouw TP, van der Hoeven JG.
    Journal: Pediatr Crit Care Med; 2009 Mar; 10(2):227-33. PubMed ID: 19057433.
    Abstract:
    OBJECTIVE: Measurement of extravascular lung water (EVLW) may be useful in the treatment of critically ill children and can be performed at the bedside using the transpulmonary thermodilution technique (TPTD). There are currently no data to verify the accuracy of these measurements in (small) children. We compared the results of TPTD measurement with the clinical gold standard transpulmonary double indicator dilution (TPDD) measurement in young children. DESIGN: Prospective clinical study in children. SETTING: Catheterization laboratory of a university hospital. PATIENTS AND METHODS: Twelve children (<2 yrs or <12 kg) under general anesthesia. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Measurements were performed using injections of ice-cold indicator (saline or dye) through a central venous catheter. Mean cardiac index was 3.91 L/min/m, mean intrathoracic blood volume index (ITBVITPDD) was 614.9 mL/m, and mean extravascular lung water index (EVLWITPDD) was 11.7 mL/kg. The correlation coefficient between EVLWITPDD and EVLWITPTD is 0.96 (95% confidence interval: 0.87-0.99; p < 0.0001). Bland-Altman analysis for EVLW measurements showed a mean bias of 2.34 mL/kg (18.13%) and limits of agreement +/-2.97 mL/kg (19.78%). The difference between measurements via the right atrium compared with the femoral vein was 2.8% for cardiac output, 8.2% for global end-diastolic volume, and 0.1% for EVLW. CONCLUSION: Clinical measurement of EVLW in young children can be performed using the TPTD with the injection catheter inserted in the femoral vein. Further studies are needed to clarify the clinical value of these measurements.
    [Abstract] [Full Text] [Related] [New Search]