These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Current practices in antinuclear antibody testing: results from the Belgian External Quality Assessment Scheme. Author: Van Blerk M, Van Campenhout C, Bossuyt X, Duchateau J, Humbel R, Servais G, Tomasi JP, Albert A, Coucke W, Libeer JC. Journal: Clin Chem Lab Med; 2009; 47(1):102-8. PubMed ID: 19072028. Abstract: BACKGROUND: This study aimed to assess the state-of-the-art of antinuclear antibody (ANA) testing as practiced in the Belgian and Luxembourg laboratories, using the results obtained in the Belgian National External Quality Assessment Scheme from 2000 to 2005. METHODS: During this period, nine samples with different specificities were sent for analysis. Participants were surveyed for methodology used and were asked to report staining pattern and titer of ANAs. In 2002, an attempt was made to improve the comparability of quantitative ANA results by the provision of a commercial reference material and to relate observed differences to methodology. RESULTS: With one exception, all participants employed a microscope-based indirect immunofluorescence assay with human epithelial cell line 2 cells. Most laboratories were accurate in describing the pattern. The percentage of unacceptable answers was greater for samples with borderline levels of antibody and for samples showing a cytoplasmic pattern. An improvement in the detection of anticentromere antibodies was observed. For all samples, a wide range of titers was reported. The provision of the secondary reference preparation led to improved inter-laboratory concordance. Comparison of methodology variables revealed a correlation between unstandardized titers and the power of the lamp of the microscope and the use of a dark room. CONCLUSIONS: The EQAS results presented in this work provide valuable insights into the state of the art of ANA testing as practiced in the Belgian and Luxembourg Laboratories and illustrate the important value of a national EQAS for ANA testing as a tool to improve performance and interlaboratory comparability of laboratory results.[Abstract] [Full Text] [Related] [New Search]