These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Comparison between Coulter LH-750 and URIT-12 for measuring hemoglobin concentration]. Author: Qin X, Liu YP, Lin FQ, Yi Z, Xu MH, Li RL, Li TJ, Li S. Journal: Nan Fang Yi Ke Da Xue Xue Bao; 2008 Dec; 28(12):2196-8. PubMed ID: 19114355. Abstract: OBJECTIVE: To evaluate the feasibility of URIT-12 hemoglobin analyzer for fast measurement of hemoglobin concentration. METHODS: Hemoglobin concentration was detected in 100 random blood samples using URIT-12 hemoglobin analyzer and Coulter LH-750 hematology analyzer. RESULTS: The two analyzers showed good correlation of the results (r=0.994) without significant difference between them (P>0.05). The linear range of URIT-12 hemoglobin analyzer was 46-240 g/L, and in the repeated measurements (20 times) of 3 batches of blood samples with low, moderate and high hemoglobin concentrations, the within-batch coefficient of variation of URIT-12 hemoglobin analyzer, from low to high concentrations, were 2.13%, 2.17%, and 2.33%, respectively. In the measurement of 4 batches of high-fat, high-bilirubin, high-globin and high-white-blood-cell blood samples, the interference rate of the former 3 samples were all less than 4% by the two devices, but that of the fourth sample was 10% by URIT-12 hemoglobin analyzer and 7% by Coulter LH-750 analyzer. CONCLUSION: The results detected by URIT-12 hemoglobin analyzer have high accuracy and precision and is easy to operate, fast-testing and portable.[Abstract] [Full Text] [Related] [New Search]