  • Title: Reduced compression mammography to reduce breast pain.
    Author: Chida K, Komatsu Y, Sai M, Nakagami A, Yamada T, Yamashita T, Mori I, Ishibashi T, Maruoka S, Zuguchi M.
    Journal: Clin Imaging; 2009; 33(1):7-10. PubMed ID: 19135922.
    Abstract:
    This study evaluated whether reduced compression mammography to relieve breast tenderness is feasible. Women can better tolerate a compression force of approximately 90 N in mammography. The breast thickness increased approximately 3 mm when the compression force was reduced to 90 N, and although the radiation dose increased approximately 20%, the image quality was identical to that with standard compression. Many patients experience breast pain with a compression force of 120 N. Reduced compression force mammography is acceptable in women whose breasts are particularly sensitive.