These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [The Pulsair manual noncontact tonometer]. Author: Bartos D, Skopový M. Journal: Cesk Oftalmol; 1991 Feb; 47(1):51-8. PubMed ID: 1913887. Abstract: The authors present a report on their own experience with a manual contact-free tonometer Pulsair of Keeler Co. Ais compared with results obtained by means of the manual aplanation Perkins tonometer of Clement Clarke International Ltd. in the group of patients with IOP under 2,8 kPa the arithmetic means of four readings, the scatter of mean values and standard deviations assessed by both methods were not significant. Correlation coefficient 0.92 (p less than 0.001). In the group of patients with glaucoma and hypertonics with IOP of 2.93 kPa and more the arithmetic means of four readings were significantly higher (at the 1% level) in contact-free tonometry. Correlation coefficient 0.95 (p less than 0.001). The authors emphasize the main advantage of the contact-free manual tonometer: safety, rapid procedure, objectivity, accuracy, manual use.[Abstract] [Full Text] [Related] [New Search]