These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Multislice radio-frequency current density imaging.
    Author: Wang D, DeMonte TP, Ma W, Joy ML, Nachman AI.
    Journal: IEEE Trans Med Imaging; 2009 Jul; 28(7):1083-92. PubMed ID: 19150780.
    Abstract:
    Radio-frequency current density imaging (RF-CDI) is an imaging technique that noninvasively measures current density distribution at the Larmor frequency utilizing magnetic resonance imaging (MRI). Previously implemented RF-CDI techniques were only able to image a single slice transverse to the static magnetic field B(0) . This paper describes the first realization of a multislice RF-CDI sequence on a 1.5 T clinical imager. Multislice RF current density images have been reconstructed for two phantoms. The influence of MRI random noise on the sensitivity of the multislice RF-CDI measurement has also been studied by theoretical analysis, simulation and phantom experiments.
    [Abstract] [Full Text] [Related] [New Search]