These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Correction of inhomogeneities in Magnetic Resonance Images. Author: Noterdaeme O, Brady M. Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2008; 2008():2221-4. PubMed ID: 19163140. Abstract: This paper presents a method by which two fast spoiled gradient echo image volumes can be used to estimate the slowly varying intensity variation frequently observed in Magnetic Resonance Images (MRI). We first present results for simulated brain MRI, where we compare the performance of a segmentation algorithm on uncorrected, corrected and artifact free images. Next, we present the results for a real uniform phantom and then for clinical spinal images, where our algorithm reduces the intensity variation within a tissue class from a factor of 30 to a factor of 2. The technique and algorithm has been successfully tested on over 1000 images acquired with different coils, sequences and anatomies.[Abstract] [Full Text] [Related] [New Search]