  • Title: Automated separation and quantitation of lipid fractions by high-performance liquid chromatography and mass detection.
    Author: Redden PR, Huang YS.
    Journal: J Chromatogr; 1991 Jun 14; 567(1):21-7. PubMed ID: 1918247.
    Abstract:
    This report describes an improved separation and quantitation of lipid fractions in a total lipid extract by high-performance liquid chromatography using a modified solvent and gradient system delivered by dual pumps and incorporating a mass detector and autosampler. The detector responses for various lipid fractions (cholesteryl esters, triacylglycerols, free cholesterol, and seven major phospholipid classes) were fitted to a quadratic equation, y = ax2 + bx + c, and quantified after detector calibration by a computer. This new system has the advantage of automation and reproducible separation. The present method was applied to rat liver analysis.